MALCOM IVC-33
summary
Characteristics of MALCOM IVC-33 viscosity controller:
- Coaxial double cylinder type
- Because the sliding speed and sliding time are certain, they have correct reproducibility.
- The spiral linear sliding pump sensor maintains a constant flow and can get a stable value.
- The viscosity of the viscous body that changes according to the stirring viscosity can be measured correctly.
- According to JIS Z8803
- The sensor is selected according to the viscosity value.
parameter
Specification parameters of MALCOM IVC-33 viscosity controller:
cylinder |
A |
B |
C |
Viscosity measurement range |
5~800Pa·S (50~8000P) |
0.20~19.99Pa·S (2.0~199.9P) |
20~1999mPa·S (20~1999CP) |
Cylinder rotation number N |
10rpm |
40rpm |
|
sliding velocity D |
6.0s-1 |
48s-1 |
192s-1 |
Sampling quantity |
5cc以上 |
40cc以上 |
60cc以上 |
Determination accuracy |
indicated value±5% |
||
repeatability |
±2% |
||
thermometry |
0~100℃FS,resolve0.1℃,precision±0.5%FS |
||
digital representation |
With viscosity·Temperature (key switch switching) |
||
Recording meter output |
viscosity・temperature |
||
Upper and lower limit output | Relay tangent point | Radio tangent point (A tangent point) resists load.120v 5A,250v 1A | |
Manual operation | Push button upper limit relay(ON-OFF Action) | ||
consumed power |
10W |
||
rectify | According to JIS Z8809 standard solution for viscosity correction and optional standard correction solution KF96. | ||
Power Supply |
AC100V,AC115V,AC220V,AC240V |
IVC-33 Viscosity Controller Accessories:
Special ultrasonic cleaner |
VS-25 |
Standard calibration solution | Silica gel KF96 100CC and 300CC |
Spare cylinder |
outer cylinder:ROM-1(A、B、C) Inner cylinder:RIM-1(A、B、C) |